DeviceSignal · Recall classified

Pride Mobility Products Corp.

FDA classified a device recall on 2026-07-31 · Class II · Z-2845-2026

EventRecall classified
Effective2026-07-31
ClassificationClass II

On 2026-07-31 the FDA classified a device recall from Pride Mobility Products Corp. as Class II. Product: Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD…. Recall Z-2845-2026. The firm initiated it on 2026-06-25. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2845-2026
Classification
Class II
Classified on
2026-07-31
Initiated on
2026-06-25
Firm
Pride Mobility Products Corp.
Reason
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Product
Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2845-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
2771d16e4a5561fa7eb3874ceba8bf659ac890aef5c0f7d60a45f8c3b8ef3aeb
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)