DeviceSignal · Recall classified

Lutronic Corporation

FDA classified a device recall on 2026-09-04 · Class II · Z-3110-2026

EventRecall classified
Effective2026-09-04
ClassificationClass II

On 2026-09-04 the FDA classified a device recall from Lutronic Corporation as Class II. Product: Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device…. Recall Z-3110-2026. The firm initiated it on 2026-08-06. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-3110-2026
Classification
Class II
Classified on
2026-09-04
Initiated on
2026-08-06
Firm
Lutronic Corporation
Reason
Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Product
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-3110-2026
Snapshot hash
e8e634c086f9f5f52e82a7eef6331d8800209bbb8cbb79bd654d29f1222caae4
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-23T12:43:31.676Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
289dcb06e2c81aac709d600e3c6c77b74af9108442a84e67db4ce7b8c08c88cc
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)