DeviceSignal · Recall classified

Boston Scientific Neuromodulation Corporation

FDA classified a device recall on 2026-08-05 · Class I · Z-2879-2026

EventRecall classified
Effective2026-08-05
ClassificationClass I

On 2026-08-05 the FDA classified a device recall from Boston Scientific Neuromodulation Corporation as Class I. Product: INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX…. Recall Z-2879-2026. The firm initiated it on 2026-06-17. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2879-2026
Classification
Class I
Classified on
2026-08-05
Initiated on
2026-06-17
Firm
Boston Scientific Neuromodulation Corporation
Reason
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
Product
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2879-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
2ead78dbd169238ac55e72ded77d6af5a1e4da2f1c8c79282010861808f16b63
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)