DeviceSignal · Recall classified

Midmark Corporation

FDA classified a device recall on 2026-08-25 · Class II · Z-3013-2026

EventRecall classified
Effective2026-08-25
ClassificationClass II

On 2026-08-25 the FDA classified a device recall from Midmark Corporation as Class II. Product: Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version…. Recall Z-3013-2026. The firm initiated it on 2026-07-20. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-3013-2026
Classification
Class II
Classified on
2026-08-25
Initiated on
2026-07-20
Firm
Midmark Corporation
Reason
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
Product
Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-3013-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
3dab864eb238c3aafec0c0784ddb0bad06cb2976b54a19794edcdd86e34a6264
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)