DeviceSignal · Recall classified

Instrumentation Laboratory

FDA classified a device recall on 2026-08-06 · Class II · Z-2899-2026

EventRecall classified
Effective2026-08-06
ClassificationClass II

On 2026-08-06 the FDA classified a device recall from Instrumentation Laboratory as Class II. Product: GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004…. Recall Z-2899-2026. The firm initiated it on 2026-05-21. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2899-2026
Classification
Class II
Classified on
2026-08-06
Initiated on
2026-05-21
Firm
Instrumentation Laboratory
Reason
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Product
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2899-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
491dcf4511b70359a8800c973434fbb61ffe541573b26dfd3528cdc0344112ac
Signal type
device_recall.classified