DeviceSignal · Recall classified

ResMed Ltd.

FDA classified a device recall on 2026-07-28 · Class I · Z-2735-2026

EventRecall classified
Effective2026-07-28
ClassificationClass I

On 2026-07-28 the FDA classified a device recall from ResMed Ltd. as Class I. Product: Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides…. Recall Z-2735-2026. The firm initiated it on 2026-06-25. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2735-2026
Classification
Class I
Classified on
2026-07-28
Initiated on
2026-06-25
Firm
ResMed Ltd.
Reason
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Product
Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2735-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
62176975636704bd20d302f92070dc51bb77be9b29cd443b7dda9580bbba3fbd
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)