DeviceSignal · Recall classified

GE Medical Systems Information Technologies Inc

FDA classified a device recall on 2026-08-04 · Class II · Z-2876-2026

EventRecall classified
Effective2026-08-04
ClassificationClass II

On 2026-08-04 the FDA classified a device recall from GE Medical Systems Information Technologies Inc as Class II. Product: GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph. Recall Z-2876-2026. The firm initiated it on 2026-06-29. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2876-2026
Classification
Class II
Classified on
2026-08-04
Initiated on
2026-06-29
Firm
GE Medical Systems Information Technologies Inc
Reason
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Product
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2876-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
80547d96309e6c74418f12caee9c1bb8384ca8f3f23537e776cb76921ffd5393
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)