DeviceSignal · Recall classified

Fresenius Medical Care Renal Therapies Group, LLC

FDA classified a device recall on 2026-09-02 · Class II · Z-3085-2026

EventRecall classified
Effective2026-09-02
ClassificationClass II

On 2026-09-02 the FDA classified a device recall from Fresenius Medical Care Renal Therapies Group, LLC as Class II. Product: 5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3. Recall Z-3085-2026. The firm initiated it on 2026-08-10. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-3085-2026
Classification
Class II
Classified on
2026-09-02
Initiated on
2026-08-10
Firm
Fresenius Medical Care Renal Therapies Group, LLC
Reason
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Product
5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-3085-2026
Snapshot hash
2ad9ab6cbee6c5eb2f0d4298731a8c9755893d917528d78bf00a015fa69a0494
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-15T12:38:10.708Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
a15a03c25f9aa749d59cdc9472adf794d7448ad79e596c590f55a5079e06ff00
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)