DeviceSignal · Recall classified

Canon Medical System, USA, INC.

FDA classified a device recall on 2026-08-25 · Class II · Z-3014-2026

EventRecall classified
Effective2026-08-25
ClassificationClass II

On 2026-08-25 the FDA classified a device recall from Canon Medical System, USA, INC. as Class II. Product: Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital…. Recall Z-3014-2026. The firm initiated it on 2026-08-04. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-3014-2026
Classification
Class II
Classified on
2026-08-25
Initiated on
2026-08-04
Firm
Canon Medical System, USA, INC.
Reason
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Product
Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-3014-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
a46e656534ec7a09b6bb559d051a2ac078be4c77b846e28a0a15ed00139ca6b9
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is