DeviceSignal · Recall classified

GE Medical Systems Information Technologies Inc

FDA classified a device recall on 2026-08-27 · Class II · Z-3036-2026

EventRecall classified
Effective2026-08-27
ClassificationClass II

On 2026-08-27 the FDA classified a device recall from GE Medical Systems Information Technologies Inc as Class II. Product: PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209. Recall Z-3036-2026. The firm initiated it on 2026-07-20. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-3036-2026
Classification
Class II
Classified on
2026-08-27
Initiated on
2026-07-20
Firm
GE Medical Systems Information Technologies Inc
Reason
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
Product
PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-3036-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
c8410a1cca4bb19e6db6f6ea642945421bc7108578cd07a65160256ad6769837
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)