DeviceSignal · Recall classified

Siemens Healthcare Diagnostics, Inc.

FDA classified a device recall on 2026-08-04 · Class III · Z-2880-2026

EventRecall classified
Effective2026-08-04
ClassificationClass III

On 2026-08-04 the FDA classified a device recall from Siemens Healthcare Diagnostics, Inc. as Class III. Product: Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No…. Recall Z-2880-2026. The firm initiated it on 2026-06-26. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2880-2026
Classification
Class III
Classified on
2026-08-04
Initiated on
2026-06-26
Firm
Siemens Healthcare Diagnostics, Inc.
Reason
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Product
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2880-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
c8987052cc937be5c55d22a9e8c496be3b4773766835bef76351b27533ab447b
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)