DeviceSignal · Recall classified

Pride Mobility Products Corp.

FDA classified a device recall on 2026-07-31 · Class II · Z-2852-2026

EventRecall classified
Effective2026-07-31
ClassificationClass II

On 2026-07-31 the FDA classified a device recall from Pride Mobility Products Corp. as Class II. Product: Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software…. Recall Z-2852-2026. The firm initiated it on 2026-06-25. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2852-2026
Classification
Class II
Classified on
2026-07-31
Initiated on
2026-06-25
Firm
Pride Mobility Products Corp.
Reason
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Product
Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2852-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
cd1cd81d3b374aef2e095dfa8e9f8ab72064b39a0556e712239aec702fd4327f
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)