DeviceSignal · Recall classified

IMRIS Imaging Inc

FDA classified a device recall on 2026-08-26 · Class II · Z-3034-2026

EventRecall classified
Effective2026-08-26
ClassificationClass II

On 2026-08-26 the FDA classified a device recall from IMRIS Imaging Inc as Class II. Product: ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000…. Recall Z-3034-2026. The firm initiated it on 2026-08-06. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-3034-2026
Classification
Class II
Classified on
2026-08-26
Initiated on
2026-08-06
Firm
IMRIS Imaging Inc
Reason
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Product
ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-3034-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
cea7746b766be04a0d2cdf15d99c58a058296a69ebe4519e0549972b599d3a14
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)