DeviceSignal · Recall classified

Beckman Coulter Inc.

FDA classified a device recall on 2026-08-17 · Class II · Z-2973-2026

EventRecall classified
Effective2026-08-17
ClassificationClass II

On 2026-08-17 the FDA classified a device recall from Beckman Coulter Inc. as Class II. Product: Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower…. Recall Z-2973-2026. The firm initiated it on 2026-07-15. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2973-2026
Classification
Class II
Classified on
2026-08-17
Initiated on
2026-07-15
Firm
Beckman Coulter Inc.
Reason
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Product
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2973-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
d305abbb8ed3dec0a914915a25494e81f224f4ad68e141510a02aa2f990d0504
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)