DeviceSignal · Recall classified

OBP Medical Corporation

FDA classified a device recall on 2026-08-10 · Class II · Z-2933-2026

EventRecall classified
Effective2026-08-10
ClassificationClass II

On 2026-08-10 the FDA classified a device recall from OBP Medical Corporation as Class II. Product: OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF…. Recall Z-2933-2026. The firm initiated it on 2026-07-13. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2933-2026
Classification
Class II
Classified on
2026-08-10
Initiated on
2026-07-13
Firm
OBP Medical Corporation
Reason
There is a potential for vaginal speculums to contain a burr on the proximal plastic blade, which may introduce a sharp edge in the contact regions for both the patient and the healthcare provider.
Product
OBP Medical , ER-Spec Single-Use Vaginal Speculum with LED Light Source, X-Small REF: C020130-1

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2933-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
e1e9aa89617815dfed31d58d4e2d4f83065d558a8982c383833f069d2c71b695
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is