DeviceSignal · Recall classified

Medtronic Perfusion Systems

FDA classified a device recall on 2026-07-27 · Class II · Z-2811-2026

EventRecall classified
Effective2026-07-27
ClassificationClass II

On 2026-07-27 the FDA classified a device recall from Medtronic Perfusion Systems as Class II. Product: EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an…. Recall Z-2811-2026. The firm initiated it on 2026-06-24. This reports the classification FDA published on that date.

Reported facts

Recall number
Z-2811-2026
Classification
Class II
Classified on
2026-07-27
Initiated on
2026-06-24
Firm
Medtronic Perfusion Systems
Reason
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Product
EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322

Every field above is as fda_device_enforcement published it. Nothing on this page is our interpretation.

Provenance

Source
fda_device_enforcement
Source record
Z-2811-2026
Snapshot hash
ab5ada73bf212eb52a548aab0968a0fcee42807c61b0350e60e24bb6e2beed72
Source URL
https://download.open.fda.gov/device/enforcement/device-enforcement-0001-of-0001.json.zip
Observed
2026-09-08T16:22:09.483Z
Transform
fda_device_enforcement_to_recall v1
Event rule
fda_device_recall_classified v1
Event key
fffc0053a9165c9774b0b52efd880a35e2862c6d523cab899db29cea6342dd7f
Signal type
device_recall.classified

Source terms that apply

More DeviceSignal signals What DeviceSignal is

Data provided by the U.S. Food and Drug Administration (https://open.fda.gov)